PENSKY MARTIEN FLASH & FIRE POINT - AUTOMATIC

Code: 11011-04


COMPLIANCE


IS:1448, ASTM D93, D3934, D3941, D3278, D3828, IP 304, 404, BS3900,6664, ISO-1516,1523,2719, DIN51 758.

APPLICATION



This apparatus is used for determination of flash points of liquid samples,whose flash points are above 450C to 3500C in closed test conditions.

 







 


For accurate and manual error free determination of flash points of fuels,


lubricating oils, and homogenous liquids as per ASTMD93, method A & B.


Test parameters are set via keypad, with digital display of set and process


temperature.


11011

We are also available on Whatsapp between 11 am to 6 pm

DESCRIPTION AND CONSTRUCTION


This apparatus is used for determination of flash points of liquid samples,whose flash points are above 450C to 3500C in closed test conditions.Outer body made of mild steel powder coated, Complete with cup made of brass, lid assembly and LPG regulator with Y-cock. There are four variants.

P.Code     MODELS         DESCRIPTION


11011-00          PFP-E                Energy Regulator Control


11011-01          PFP-V                Voltage Variac Control


11011-02          PFP-SA              Semi Automatic


11011-03          PFP-AG              Automatic


11011-04          PFP-AE              Automatic with Electric Ignition


11011-05          PFP-PRIME        Automatic Prime

AUTOMATIC MODEL


For accurate and manual error free determination of flash points of fuels,lubricating oils, and homogenous liquids as per ASTMD93, method A & B.Test parameters are set via keypad, with digital display of set and process temperature.

STANDARD FEATURES FOR AUTOMATIC


* Automatic Flash point detection


* Automatic Barometric pressure correction


* Automatic flash point search mode for unknown samples.


* User set-able heating and flashing profile


* Automatic or user set-able stirrer speed setting and control


* Automatic heating profile setting depending on initial sample temperature and expected flash point.


* 16 x 4 Line Back Lit Alpha numeric display


* Automatic cutoff at 30°C beyond expected flash point or at max. temperature limit set by the user.


* 2 Preset test Modes as per ASTM Method A & B


* In-bulit Numeric Key pad for settings


* Audio visual indication of end of test and Alarms


* Over temperature safety cutoff and alarm.


* Last reading stored until report is printed or next test is initiated.


* Serial Printer interface for Automatic printing of test report at a press of a button.


* Ensures min. power consumption by design thereby reducing operation cost.


* Electrical ignition instead of gas (Optional)

ACCESSORIES / SPARES


11011-11       Energy Regulator


11011-12       Voltage Variac


11011-13       Digital Controller


11011-14       Reducing socket cum Flame regulator


11011-15       IP-15C Thermometer


11011-16       IP-16C Thermometer


11011-16       Brass Test cup


11011-17       Lid with Shutter


11011-18       Open Cup Lid


11011-19       Motorized Stirrer (02/03 Only)


11011-20       Heating Coil


11011-21       Gas Tubing


11011-22       Temperature Sensor


xxxx-xx          Portable digital barometer


11011-23        Spare electrical ignition coil. (Automatic Models)


11011-24        Lid with shutter for electrical ignition(Automaticmodels) 


11011-25        Spare flash detector (Automatic Models)


11011-26        Calibration Certification